Pulse Transit Time Derivation Using Xiphoidal and Carotid Seismocardiograms PUBLIC ACCESS

[+] Author Affiliations
Ajay K. Verma, John Zanetti, Reza Fazel-Rezai, Kouhyar Tavakolian

University of North Dakota

Paper No. DMD2017-3444, pp. V001T01A010; 2 pages
doi:10.1115/DMD2017-3444
From:
  • 2017 Design of Medical Devices Conference
  • 2017 Design of Medical Devices Conference
  • Minneapolis, Minnesota, USA, April 10–13, 2017
  • ISBN: 978-0-7918-4067-2
  • Copyright © 2017 by ASME

abstract

Blood pressure is an indicator of a cardiovascular functioning and could provide early symptoms of cardiovascular system impairment. Blood pressure measurement using catheterization technique is considered the gold standard for blood pressure measurement [1]. However, due its invasive nature and complexity, non-invasive techniques of blood pressure estimation such as auscultation, oscillometry, and volume clamping have gained wide popularity [1]. While these non-invasive cuff based methodologies provide a good estimate of blood pressure, they are limited by their inability to provide a continuous estimate of blood pressure [1–2]. Continuous blood pressure estimate is critical for monitoring cardiovascular diseases such as hypertension and heart failure.

Pulse transit time (PTT) is a time taken by a pulse wave to travel between a proximal and distal arterial site [3]. The speed at which pulse wave travels in the artery has been found to be proportional to blood pressure [1, 3]. A rise in blood pressure would cause blood vessels to increase in diameter resulting in a stiffer arterial wall and shorter PTT [1–3]. To avail such relationship with blood pressure, PTT has been extensively used as a marker of arterial elasticity and a non-invasive surrogate for arterial blood pressure estimation.

Typically, a combination of electrocardiogram (ECG) and photoplethysmogram (PPG) or arterial blood pressure (ABP) signal is used for the purpose of blood pressure estimation [3], where the proximal and distal timing of PTT (also referred as pulse arrival time, PAT) is marked by R peak of ECG and a foot/peak of a PPG, respectively. In the literature, it has been shown that PAT derived using ECG-PPG combination infers an inaccurate estimate of blood pressure due to the inclusion of isovolumetric contraction period [1–3, 4].

Seismocardiogram (SCG) is a recording of chest acceleration due to heart movement, from which the opening and closing of the aortic valve can be obtained [5]. There is a distinct point on the dorso-ventral SCG signal that marks the opening of the aortic valve (annotated as AO). In the literature, AO has been proposed for timing the onset of the proximal pulse of the wave [6–8]. A combination of AO as a proximal pulse and PPG as a distal pulse has been used to derive pulse transit time and is shown to be correlated with blood pressure [7]. Ballistocardiogram (BCG) which is a measure of recoil forces of a human body in response to pumping of blood in blood vessels has also been explored as an alternative to ECG for timing proximal pulse [5, 9].

Use of SCG or BCG for timing the proximal point of a pulse can overcome the limitation of ECG-based PTT computation [6–7, 9]. However, a limitation of current blood pressure estimation systems is the requirement of two morphologically different signals, one for annotating the proximal (ECG, SCG, BCG) and other for annotating the distal (PPG, ABP) timing of a pulse wave. In the current research, we introduce a methodology to derive PTT from seismocardiograms alone. Two accelerometers were used for such purpose, one was placed on the xiphoid process of the sternum (marks proximal timing) and the other one was placed on the external carotid artery (marks distal timing). PTT was derived as a time taken by a pulse wave to travel between AO of both the xiphoidal and carotid SCG.

Copyright © 2017 by ASME
This article is only available in the PDF format.

Figures

Tables

Interactive Graphics

Video

Country-Specific Mortality and Growth Failure in Infancy and Yound Children and Association With Material Stature

Use interactive graphics and maps to view and sort country-specific infant and early dhildhood mortality and growth failure data and their association with maternal

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In