Prediction of Shear Induced Platelet Activation in Prosthetic Heart Valves by Integrating Fluid–Structure Interaction Approach and Lagrangian-Based Blood Damage Model

[+] Author Affiliations
Umberto Morbiducci, Diana Massai, Franco M. Montevecchi

Politecnico di Torino, Turin, Italy

Raffaele Ponzini

CILEA, Milan, Italy

Matteo Nobili, Alberto Redaelli

Politecnico di Milano, Milan, Italy

Danny Bluestein

Stony Brook University, Stony Brook, NY

Paper No. SBC2009-206162, pp. 63-64; 2 pages
doi:10.1115/SBC2009-206162
From:
  • ASME 2009 Summer Bioengineering Conference
  • ASME 2009 Summer Bioengineering Conference, Parts A and B
  • Lake Tahoe, California, USA, June 17–21, 2009
  • Conference Sponsors: Bioengineering Division
  • ISBN: 978-0-7918-4891-3
  • Copyright © 2009 by ASME

abstract

Altered haemodynamics are implicated in the blood cells damage that leads to thromboembolic complications in presence of prosthetic cardiovascular devices, with platelet activation being the underlying mechanism for cardioemboli formation in blood flow past mechanical heart valves (MHVs). Platelet activation can be initiated and maintained by flow patterns arising from blood flowing through the MHV, and can lead to an enhancement in the aggregation of platelets, increasing the risk for thromboemboli formation. Hellums and colleagues compiled numerous experimental results to depict a locus of incipient shear related platelet activation on a shear stress – exposure time plane, commonly used as a standard for platelet activation threshold [1]. However, platelet activation and aggregation is significantly greater under pulsatile or dynamic condition relative to exposure to constant shear stress [2]. Previous studies do not allow to determine the relationship existing between the measured effect — the activation of a platelet, and the cause — the time-varying mechanical loading, and the time of exposure to it as might be expected in vivo when blood flows through the valve. The optimization of the thrombogenic performance of MHVs could be facilitated by formulating a robust numerical methodology with predictive capabilities of flow-induced platelet activation. To achieve this objective, it is essential (i) to quantify the link between realistic valve induced haemodynamics and platelet activation, and (ii) to integrate theoretical, numerical, and experimental approaches that allow for the estimation of the thrombogenic risk associated with a specific geometry and/or working conditions of the implantable device. In this work, a comprehensive analysis of the Lagrangian systolic dynamics of platelet trajectories and their shear histories in the flow through a bileaflet MHV is presented. This study uses information extracted from the numerical simulations performed to resolve the flow field through a realistic model of MHV by means of an experimentally validated fluid-structure interaction approach [3]. The potency of the device to mechanically induce activation/damage of platelets is evaluated using a Lagrangian-based blood damage cumulative model recently identified using in vitro platelet activity measurements [4,5].

Copyright © 2009 by ASME

Figures

Tables

Interactive Graphics

Video

Country-Specific Mortality and Growth Failure in Infancy and Yound Children and Association With Material Stature

Use interactive graphics and maps to view and sort country-specific infant and early dhildhood mortality and growth failure data and their association with maternal

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In